Abstract
Mobile eye-tracking is important for understanding the role of vision during real-world tasks in older adults (OA) and people with Parkinson's disease (PD). However, accuracy and reliability of such devices have not been established in these populations. We used a novel protocol to quantify accuracy and reliability of a mobile eye-tracker in OA and PD.A mobile eye-tracker (Dikablis) measured the saccade amplitudes of 20 OA and 14 PD on two occasions. Participants made saccades between targets placed 5°, 10° and 15° apart. Impact of visual correction (glasses) on saccadic amplitude measurement was also investigated in 10 OA.Saccade amplitude accuracy (median bias) was -1.21° but a wide range of bias (-7.73° to 5.81°) was seen in OA and PD, with large vertical saccades (15°) being least accurate. Reliability assessment showed a median difference between sessions of <1° for both groups, with poor to good relative agreement (Spearman rho: 0.14 to 0.85). Greater accuracy and reliability was observed in people without visual correction.Saccade amplitude can be measured with variable accuracy and reliability using a mobile eye-tracker in OA and PD. Human, technological and study-specific protocol factors may introduce error and are discussed along with methodological recommendations.
Original language | English |
---|---|
Pages (from-to) | 308-315 |
Number of pages | 8 |
Journal | Medical Engineering and Physics |
Volume | 38 |
Issue number | 3 |
Early online date | 16 Jan 2016 |
DOIs | |
Publication status | Published - 1 Mar 2016 |
Externally published | Yes |