Abstract
A mid-infrared tunable diode laser absorption spectroscopy (TDLAS) gas sensor based on hollow waveguide (HWG) gas cell for real-time exhaled carbon dioxide (eCO2) detection is reported. A 2.73 μm distributed feedback (DFB) laser was used to target a strong CO2 absorption line, and wavelength modulation spectroscopy (WMS) with the second harmonic (WMS-2.) was used to retrieve the CO2 concentration with high sensitivity. The influence of different parameters, including coupling configuration of HWG, laser-to-HWG and HWG-to-detector coupling alignment on the stability of the HWG sensor is systematically studied. The HWG eCO2 sensor showed a fast response time of 2.7s, detection limit of 17 ppmv, and measurement precision of 20.9 ppmv with a 0.54 s temporal resolution. The eCO2 concentrations changed in breath cycles were measured in real time. The Allan variance indicated that the detection limit can reach 1.7 ppmv, corresponding to a detection sensitivity of 1.3(215)10-8 cm-1Hz-1/2, as the integration time increases to 26 s. This work demonstrates the performance characteristics and merits of HWG eCO2 sensor for exhaled breath analysis and potential detection for other exhaled gases.
| Original language | English |
|---|---|
| Article number | 8825859 |
| Pages (from-to) | 11972-11979 |
| Number of pages | 8 |
| Journal | IEEE Sensors Journal |
| Volume | 19 |
| Issue number | 24 |
| Early online date | 5 Sept 2019 |
| DOIs | |
| Publication status | Published - 15 Dec 2019 |
Keywords
- Tunable diode laser absorption spectroscopy
- carbon dioxide
- distributed feedback laser
- hollow waveguide
Fingerprint
Dive into the research topics of 'Influence of light coupling configuration and alignment on the stability of HWG-based gas sensor system for real-time detection of exhaled carbon dioxide'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver