Low-Cost QCM Sensor System for Screening Semen Samples

Michael Newton, Shaun Atherton, Robert Morris, Simon Stanley, Carl Evans, David Hughes, Glen McHale

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)
14 Downloads (Pure)

Abstract

Artificial insemination is a well-established part of modern agricultural practice. A viable semen sample is judged by the total number of spermatozoa (sperm) in the sample and the motility of the sperm. In this paper, we report the development of a reusable measurement cell and electronics for screening semen samples based on the Quartz Crystal Microbalance (QCM) and Universal Frequency to Digital Converter (UFDC-1) to produce a low-cost sensor system. After introducing the semen sample at one end of the measurement cell, sperm swim down a channel before causing a frequency change on the QCM. Data is presented that shows the different frequency changes using a commercial frequency counter caused by porcine semen samples, one two days old and one twenty one days old. Similar data is presented for a motile semen sample measurement using the low-cost UFDC-1.
Original languageEnglish
Pages (from-to)1-5
JournalJournal of Sensors
Volume2010
Issue number326365
DOIs
Publication statusPublished - 2010

Fingerprint

Dive into the research topics of 'Low-Cost QCM Sensor System for Screening Semen Samples'. Together they form a unique fingerprint.

Cite this