Sensitivity enhancement in ICP emission spectrometry with microcolumn preconcentration and ultrasonic nebulisation

Watcharapong Worrasettapong, Renli Ma, Alan G. Cox, Cameron W. McLeod*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

3 Citations (Scopus)

Abstract

To improve the measurement capability of ICP emission spectrometry, microcolumn preconcentration using an iminodiacetate chelating resin (Muromac A-1) has been combined with ultrasonic nebulisation for determination of ultratrace elements (V, Mn, Fe, Co, Ni, Cu, Zn, Cd, Pb and U) in natural waters including seawaters. Trace element deposition was achieved at pH 5 and elution was effected by injection of nitric acid (250 μl, 2.0 M). Sensitivity enhancement factors between 26 and 44 were achieved with on-line preconcentration (sample volume of 10 ml) and detection limits were improved by up to two orders of magnitude relative to conventional pneumatic nebulisation of original water samples. The method was successfully applied to mineral, rain and sea water samples.

Original languageEnglish
Pages (from-to)825-834
Number of pages10
JournalInternational Journal of Environmental Analytical Chemistry
Volume82
Issue number11-12
DOIs
Publication statusPublished - Dec 2002
Externally publishedYes

Keywords

  • Axial viewing
  • Flow injection
  • ICP emission spectrometry
  • Microcolumn preconcentration
  • Natural waters
  • Ultrasonic nebulisation

Fingerprint

Dive into the research topics of 'Sensitivity enhancement in ICP emission spectrometry with microcolumn preconcentration and ultrasonic nebulisation'. Together they form a unique fingerprint.

Cite this